Reverse Laminar Air Flow (RLAF) booth designed to provide operator protection from airborne dust particles during bulk chemical dispensing, raw material sampling, and powder weighing.
ADS Biotech Dispensing & Sampling Booths (RLAF Booths) draw air away from the operator's breathing zone into low-level return grilles. The contaminated air passes through 3 stages of filtration before being recirculated down into the booth as clean, sterile laminar air. This continuous sweeping action ensures powder containment while keeping the operator and environment safe.
| Parameter | Specification |
|---|---|
| MOC | SS 304 / SS 316 double skin structure with coved corners |
| Filtration Efficiency | Supply HEPA 99.997% @ 0.3µm, Intermediate 95%, Pre-filter 90% |
| Airflow Velocity | 90 ± 20 FPM (0.45 m/s) downward airflow |
| Curtains | Transparent PVC strip curtain / SS side panel enclosures |